Skip to main content

JUMO screw-in RTD temperature probes with form J connection head, with continuous protection tube

Product number: 802012

Product information "JUMO screw-in RTD temperature probes with form J connection head, with continuous protection tube"

Screw-in resistance thermometer with form J connection head

Type: 902030/10-380-2003-1-6-50-104/000

  •  With continuous protection tube
  •  Measuring range from -50 to +200 °C
  •  Single resistance thermometer
  •  2x Pt 100 in two-wire circuit
  •  Tolerance class: Class B (standard)
  •  Thermowell diameter: 6 mm
  •  Installation length: 50 mm
  •  Process connection: screw connection G ½
  •  Extra codes: none

The screw-in RTD temperature probe with connection head is specially designed for use in various industries such as air conditioning and refrigeration technology and heating, furnace and apparatus engineering. It offers precise temperature measurement in both liquid and gaseous media, with a temperature range of -50 to +400 °C.
The installation design of this thermometer ensures reliable tightness, even under extreme conditions of negative and positive pressure.
The measuring insert is fitted as standard with a Pt100 temperature sensor in accordance with the DIN EN 60751:2009 / IEC 60751:2008 standards. This sensor has a temperature coefficient α of 3.850 × 10-3 °C-1 and is installed in a two-wire circuit. Versions with Pt500 or Pt1000 are also available as an option. For more precise measurement results, the connection can be made either in a three-wire or four-wire circuit, and it is also possible to integrate a transmitter into the connection head.
Depending on the measurement requirements, a single or dual resistance thermometer can also be selected.

Dimensions

Height 11 cm
Length 5,9 cm
Width 4,9 cm

Weight

Weight 190 g

Other attributes

Country of origin DE
Customs tariff number 90251900
Einbaulänge in mm 50 mm
Messeinsatz 2x Pt 100 Zweileiterschaltung
Prozessanschluss Verschraubung G 1/2
Temperaturbereich -50°C ... +200°C
Technical data:

Connection head:

Form J, die-cast aluminium, M16 × 1.5; IP65, ambient temperature -40 to +100 °C Attention: reduced ambient temperature when using transmitters, data sheet 707030.

If the design permits, yellow connection heads are used for SIL and PL applications (extra codes 658 and 659). Exceptions are sensors for use in alkaline environments and sensors for which LABS freedom is mandatory.

Process connection:

Thread, stainless steel 1.4571

Thermowell:

Stainless steel 1.4571, Ø 6 mm and Ø 8 mm

Measuring insert:

Pt100 temperature sensor, DIN EN 60751:2009 / IEC 60751:2008, class B, two-wire circuit

Response times:

t0.5 = 8 s, t0.9 = 20 s, in water 0.4 m/s, Ø 8 mm

t0.5 = 5 s, t0.9 = 12 s, in water 0.4 m/s, Ø 6 mm

t0.5 = 40 s, t0.9 = 110 s, in air 3.0 m/s, Ø 6 mm

t0.5 = 2 s, t0.9 = 5 s, in water 0.4 m/s, Ø 6 mm reduced to Ø 3.5 mm

t0.5 = 25 s, t0.9 = 85 s, in air 3.0 m/s, Ø 6 mm stepped down to Ø 3.5 mm

t0.5 = 1 s, t0.9 = 2 s, in water 0.4 m/s, Ø 6 mm reduced to Ø 2.6 mm

t0.5 = 18 s, t0.9 = 65 s, in air 3.0 m/s, Ø 6 mm stepped down to Ø 2.6 mm

Measuring transducer:

analogue transmitter, output 4 to 20 mA, data sheet 707030

Accessories:

Protective sleeve, data sheet 909710