JUMO tecLine CL2 sensor for free chlorine

Product number: 802182

Product information "JUMO tecLine CL2 sensor for free chlorine"

These amperometric sensors with membrane are used to determine the concentration of free chlorine. They are able to measure various inorganic chlorine compounds, including chlorine gas (Cl₂), electrolytically generated chlorine, sodium hypochlorite (NaOCl), calcium hypochlorite (Ca(OCl)₂) and chlorinated lime (Ca(OCl)Cl).

Testing for the complete absence of free chlorine is not possible with these sensors.

Depending on the version, the sensors either supply a temperature-compensated current signal (4-20 mA) in an analogue version or a Modbus RTU signal via a digital interface. Calibration takes place in a downstream device such as an indicator, controller, recorder or programmable logic controller (PLC).

The sensors can be connected directly to various transmitters or controllers. These supply the system with the required voltage and enable uncomplicated calibration.

Areas of application:

The sensors are used to measure free chlorine in drinking water, bathing water, industrial water, process water and cooling water.

Dimensions

Height 8 cm
Length 30 cm
Width 20 cm

Weight

Weight 640 g

Other attributes

Country of origin DE
Customs tariff number 90278930
Technical data

Sensors for free chlorine

 

Sensor type

202630/40

202630/50

Measurand

Free chlorine

Area of application

Swimming pool water, drinking water, service water and process water

Suitable chlorination agents

Inorganic chlorine compounds: NaOCl (sodium hypochlorite), Ca(OCl)2, chlorine gas, chlorine pro- duced by membrane electrolysis (not suitable: chlorine electrolysis without a membrane)

Measuring principle

Membrane-covered, amperometric, two-electrode system with integrated electronics

Membrane type

Hydrophobic PTFE membrane

Measuring cable connection

2-pin terminal connection (2 × 1 mm2)

5-pin flange connector, M12

Voltage supply

UB DC 12 to 30 V

(galvanic isolation required)

UB DC 22.5 to 26 V

(galvanically isolated from the sensor)

Electromagnetic compatibilitya

Interference emission: class Bb Interference immunity: to industrial requirements

Output signal

4 to 20 mA

Modbus RTU

Load/current consumption

≤ (UB - 7.5 V) ÷ 0.02 A

approx. 20 mA

Settling time

1 h

Inflow speed

Approx. 15 cm/s

corresponds to a flow of approx. 30 l/h when installed in the JUMO flow fitting (part no.: 00392611)

Measuring rangesc

0.05 to 0.5 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 5 mg/l (ppm)

0.05 to 10 mg/l (ppm)

0.05 to 100 mg/l (ppm)

0.05 to 200 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 20 mg/l (ppm)

Resolution

0.01 mg/l with measuring range 0.5/2/5/10 mg/l

0.1 mg/l with measuring range 100/200 mg/l

0.001 mg/l with measuring range 2 mg/l

0.01 mg/l with measuring range 20 mg/l

Accuracyd

Measuring range 2 mg/l

 

< 1 % with 0,4 mg/l

< 1 % with 1,6 mg/l

Slope drifte

Approx. < -1 % per month

Response time t90

approx. 30 s

Operating temperature

Sample water temperature Ambient temperature

 

0 to 45 °Cf

0 to 55 °C

Temperature compensation

Automatic, using integrated temperature probe

Zero point adjustment

Not required

Slope adjustment

On evaluation unit/controller using analytical chlorine determination (DPD-1-method)

pH value operating range

pH 6 to pH 8

Note the effect of the pH value on the disinfecting properties, corrosion and the dissociation curve.

pH dependence (loss of slope)

with pH 8, approx. 65%

with pH 9, approx. 95% (starting at pH 7)

Disturbances

ClO2: recorded with a concentration factor of 9 O3: recorded

Chlorine electrolysis without a membrane can cause disturbances

Pressure resistanceg

Pabs max. 2 bar Prel max. 1 bar

Materials

Semi-permeable membrane, PVC-U

Dimensions

Dia. 25 mm, length 220 mm (housing with membrane cap)

Dia. 25 mm, length 205 mm (housing with membrane cap)

Weight

Approx. 125 g

a  EN 61326-1, EN 61326-2-3.

b  The product is suitable for industrial use as well as for households and small businesses.

c  Other measuring ranges upon request.

d  After calibration under replicable conditions (25 °C, pH 7.2 in drinking water) from the measuring range end value.

e  Under replicable conditions (25 °C, pH 7.2 in drinking water).

f  Prerequisite: no ice crystals in the measurement medium.

g  Pressure fluctuations are not admissible. Pressure-free operation (atmospheric pressure) recommended.

Skip product gallery

We recommend the following

JUMO single fitting flow monitor connection M12 process connection
Product number: 802153
Flow monitors for disinfection variables Membrane-covered sensors for water disinfection (e.g. free chlorine or total chlorine, ozone, chlorine dioxide, etc.) must be operated with a minimum flow rate of 15 cm/s. The flow monitor is integrated into the sample water supply line, in series with the disinfection sensor, and monitors the required minimum flow to the sensor. Below the minimum flow velocity, the sensors measure values that are too low. This can cause a connected control system to dangerously overdose or underdose. Above the minimum incident flow velocity, the measuring signal is only slightly influenced by the incident flow velocity. The flow monitor consists of a flow body, a needle valve insert for flow regulation and an inductive proximity switch, which is controlled by a float in the medium flow. When the minimum flow velocity is reached or exceeded, the float rises and the contact of the inductive proximity switch switches. The contact can be used, for example, to control a binary input of the controller for standard signals, JUMO AQUIS 500 AS. If the flow is too low, the controller is set to the ‘HOLD’ state. This prevents incorrect dosing.   Areas of application Flow monitoring when using sensors for free chlorine, chlorine dioxide, ozone, hydrogen peroxide or peracetic acid