JUMO tecLine CL2 sensor for free chlorine

Product number: 802185

Product information "JUMO tecLine CL2 sensor for free chlorine"

These amperometric sensors with membrane are used to determine the concentration of free chlorine. They are able to measure various inorganic chlorine compounds, including chlorine gas (Cl₂), electrolytically generated chlorine, sodium hypochlorite (NaOCl), calcium hypochlorite (Ca(OCl)₂) and chlorinated lime (Ca(OCl)Cl).

Testing for the complete absence of free chlorine is not possible with these sensors.

Depending on the version, the sensors either supply a temperature-compensated current signal (4-20 mA) in an analogue version or a Modbus RTU signal via a digital interface. Calibration takes place in a downstream device such as an indicator, controller, recorder or programmable logic controller (PLC).

The sensors can be connected directly to various transmitters or controllers. These supply the system with the required voltage and enable uncomplicated calibration.

Areas of application:

The sensors are used to measure free chlorine in drinking water, bathing water, industrial water, process water and cooling water.

Dimensions

Height 8 cm
Length 30 cm
Width 20 cm

Weight

Weight 640 g

Other attributes

Country of origin DE
Customs tariff number 90278930
Technical data

Sensors for free chlorine

 

Sensor type

202630/40

202630/50

Measurand

Free chlorine

Area of application

Swimming pool water, drinking water, service water and process water

Suitable chlorination agents

Inorganic chlorine compounds: NaOCl (sodium hypochlorite), Ca(OCl)2, chlorine gas, chlorine pro- duced by membrane electrolysis (not suitable: chlorine electrolysis without a membrane)

Measuring principle

Membrane-covered, amperometric, two-electrode system with integrated electronics

Membrane type

Hydrophobic PTFE membrane

Measuring cable connection

2-pin terminal connection (2 × 1 mm2)

5-pin flange connector, M12

Voltage supply

UB DC 12 to 30 V

(galvanic isolation required)

UB DC 22.5 to 26 V

(galvanically isolated from the sensor)

Electromagnetic compatibilitya

Interference emission: class Bb Interference immunity: to industrial requirements

Output signal

4 to 20 mA

Modbus RTU

Load/current consumption

≤ (UB - 7.5 V) ÷ 0.02 A

approx. 20 mA

Settling time

1 h

Inflow speed

Approx. 15 cm/s

corresponds to a flow of approx. 30 l/h when installed in the JUMO flow fitting (part no.: 00392611)

Measuring rangesc

0.05 to 0.5 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 5 mg/l (ppm)

0.05 to 10 mg/l (ppm)

0.05 to 100 mg/l (ppm)

0.05 to 200 mg/l (ppm)

0.05 to 2 mg/l (ppm)

0.05 to 20 mg/l (ppm)

Resolution

0.01 mg/l with measuring range 0.5/2/5/10 mg/l

0.1 mg/l with measuring range 100/200 mg/l

0.001 mg/l with measuring range 2 mg/l

0.01 mg/l with measuring range 20 mg/l

Accuracyd

Measuring range 2 mg/l

 

< 1 % with 0,4 mg/l

< 1 % with 1,6 mg/l

Slope drifte

Approx. < -1 % per month

Response time t90

approx. 30 s

Operating temperature

Sample water temperature Ambient temperature

 

0 to 45 °Cf

0 to 55 °C

Temperature compensation

Automatic, using integrated temperature probe

Zero point adjustment

Not required

Slope adjustment

On evaluation unit/controller using analytical chlorine determination (DPD-1-method)

pH value operating range

pH 6 to pH 8

Note the effect of the pH value on the disinfecting properties, corrosion and the dissociation curve.

pH dependence (loss of slope)

with pH 8, approx. 65%

with pH 9, approx. 95% (starting at pH 7)

Disturbances

ClO2: recorded with a concentration factor of 9 O3: recorded

Chlorine electrolysis without a membrane can cause disturbances

Pressure resistanceg

Pabs max. 2 bar Prel max. 1 bar

Materials

Semi-permeable membrane, PVC-U

Dimensions

Dia. 25 mm, length 220 mm (housing with membrane cap)

Dia. 25 mm, length 205 mm (housing with membrane cap)

Weight

Approx. 125 g

a  EN 61326-1, EN 61326-2-3.

b  The product is suitable for industrial use as well as for households and small businesses.

c  Other measuring ranges upon request.

d  After calibration under replicable conditions (25 °C, pH 7.2 in drinking water) from the measuring range end value.

e  Under replicable conditions (25 °C, pH 7.2 in drinking water).

f  Prerequisite: no ice crystals in the measurement medium.

g  Pressure fluctuations are not admissible. Pressure-free operation (atmospheric pressure) recommended.

Skip product gallery

We recommend the following

JUMO single fitting flow monitor connection M12 process connection
Product number: 802153
Flow monitors for disinfection variables Membrane-covered sensors for water disinfection (e.g. free chlorine or total chlorine, ozone, chlorine dioxide, etc.) must be operated with a minimum flow rate of 15 cm/s. The flow monitor is integrated into the sample water supply line, in series with the disinfection sensor, and monitors the required minimum flow to the sensor. Below the minimum flow velocity, the sensors measure values that are too low. This can cause a connected control system to dangerously overdose or underdose. Above the minimum incident flow velocity, the measuring signal is only slightly influenced by the incident flow velocity. The flow monitor consists of a flow body, a needle valve insert for flow regulation and an inductive proximity switch, which is controlled by a float in the medium flow. When the minimum flow velocity is reached or exceeded, the float rises and the contact of the inductive proximity switch switches. The contact can be used, for example, to control a binary input of the controller for standard signals, JUMO AQUIS 500 AS. If the flow is too low, the controller is set to the ‘HOLD’ state. This prevents incorrect dosing.   Areas of application Flow monitoring when using sensors for free chlorine, chlorine dioxide, ozone, hydrogen peroxide or peracetic acid

€461.13*
JUMO ecoLine glass shaft pH electrode with KCL container, S8 Pg13.5 screw head
Product number: 800191
Diaphragm: Ceramic | Electrical connection of the electrode (onto which the cable is plugged): Screw head Pg13.5 (S8) | Extra code: KCL container (052) | Insertion length: 120mm | Measured parameter: pH | Measuring resistor (temperature sensor): without
pH/ORP electrodes with glass shaftSelected version: pH 0 to 12, short term pH 14without temperature sensor (not available for this series, only for Jumo tecLine and tecLine HD series) Screw head S8 Pg13.5 Ceramic diaphragm Installation length 120mm (standard) KCL containerType code (can also be found on your electrode): 201005/51-18-22-0000-00-120/052 ,part no. 00402273The electrodes of the JUMO ecoLine series are high-quality measuring sensors with a favourable price/performance ratio.Active part pHJUMO ecoLine electrodes  are equipped with the low-resistance JUMO UW glass. This guarantees fast and reliable measurement results.Active part ORPA robust platinum tip provides reliable measured values and allows easy cleaning of the sensor.Reference systemA polymerised KCl solid electrolyte is used in the glass version. This is ideally suited for measurement in general aqueous media. aqueous media. For measurements in media with lower ion content or at high flow rates, the electrolyte can be the electrolyte can be equipped with an optional "salt reservoir" to increase the service life (if wished, please use our configurator). The JUMO cartridge discharge system, which has been tried and tested for years, is used as the reference system. Here, the electrolyte remains free of silver ions over the service life of the sensor. and therefore less susceptible to electrode poisons.Optionally, the electrodes can be supplied in a storage container (quiver). This is recommended for sporadic use of the sensor, for refreshing the sensor after longer, intensive use or for longer storage periods.All sensors can be supplied as OEM versions on request. Special versions on request.Areas of applicationDrinking water monitoring and treatmentSwimming pool - Aquaristics (also seawater aquaristics)Greenhouse technologyLightly polluted industrial, process and waste waterRainwater, pond water and surface waterhand-held measuring devicesNot recommended forgalvanic water treatmentindustrial process and waste waterultrapure watertherapeutic bathsbiotechnology, sterilisation processes

From €173.93*
JUMO ecoLine glass shaft pH electrode with salt reserve, S8 Pg13.5 screw head
Product number: 800192
Diaphragm: Ceramic | Electrical connection of the electrode (onto which the cable is plugged): Screw head Pg13.5 (S8) | Extra code: Salt reservoir (837) | Insertion length: 120mm | Measured parameter: pH | Measuring resistor (temperature sensor): without
pH/ORP electrodes with glass shaftSelected version: pH 0 to 12, short term pH 14without temperature sensor (not available for this series, only for Jumo tecLine and tecLine HD series) Screw head S8 Pg13.5 Ceramic diaphragm Installation length 120mm (standard) Salt reserveType code (can also be found on your electrode): 201005/51-18-22-0000-00-120/837 ,part no. 00447837 The electrodes of the JUMO ecoLine series are high-quality measuring sensors with a favourable price/performance ratio.Active part pHJUMO ecoLine electrodes  are equipped with the low-resistance JUMO UW glass. This guarantees fast and reliable measurement results.Active part ORPA robust platinum tip provides reliable measured values and allows easy cleaning of the sensor.Reference systemA polymerised KCl solid electrolyte is used in the glass version. This is ideally suited for measurement in general aqueous media. aqueous media. For measurements in media with lower ion content or at high flow rates, the electrolyte can be the electrolyte can be equipped with an optional "salt reservoir" to increase the service life (if wished, please use our configurator). The JUMO cartridge discharge system, which has been tried and tested for years, is used as the reference system. Here, the electrolyte remains free of silver ions over the service life of the sensor. and therefore less susceptible to electrode poisons.Optionally, the electrodes can be supplied in a storage container (quiver). This is recommended for sporadic use of the sensor, for refreshing the sensor after longer, intensive use or for longer storage periods.All sensors can be supplied as OEM versions on request. Special versions on request.Areas of applicationDrinking water monitoring and treatmentSwimming pool - Aquaristics (also seawater aquaristics)Greenhouse technologyLightly polluted industrial, process and waste waterRainwater, pond water and surface waterhand-held measuring devicesNot recommended forgalvanic water treatmentindustrial process and waste waterultrapure watertherapeutic bathsbiotechnology, sterilisation processes

From €173.34*
JUMO AQUIS RS, indicator/controller for digital sensors with Modbus protocol in surface-mounting housing, with RS485 Modbus master, analog output 0 (4) - 20mA or 0 - 10V output for main value
Product number: 800193
Unit selection: Version 1: Stock version: Type 202569/20-654-888-000-000-23/000 ; Part No. 00613453
Indicator/controller for digital sensors with Modbus protocolThe unit has 1 digital interface, 1 analogue temperature input and 1 binary input. The digital interface is suitable for connecting sensors that have a Modbus RS485 interface. If the connected sensor does not have an integrated temperature sensor, a resistance thermometer Pt100, Pt1000 or NTC/PTC (up to 4 k) can be connected to the analogue temperature input. The power supply for the digital sensor is integrated in the unit.The LCD graphic display allows the input signals to be shown as digits or as a bar graph. The display of the parameters in plain text makes operation easy to understand and safe.With two optional relay change-over contacts, both simple switching or alarm functions and sophisticated control tasks with P, PI, PD and PID behaviour can be realised. On request, the unit can also be supplied with two freely parameterisable and scalable analogue outputs (0 to 10 V, or 0(4) to 20 mA).Special featuresDisplay: mg/l, ppm, % SAT, %, ‰, g/l, ppb, µg/l, NTU, FNU, etc.Special displays are also possible with the setup programmeLarge LCD graphic display with backlightingDisplay representation selectable: large digits, bargraph or trend displayIntegrated calibration routines:depending on the sensor: zero point, full scale and 2-pointCalibration logbookProtection class IP67 for surface mountingProtection class IP65 for control cabinet mountingOperator languages (switchable): German, English, Frenchthrough setup program (option): convenient programming, system documentation, reloading of additional operating languages

€866.80*