Product information "JUMO wall resistance thermometer with outdoor sensor Protection class IP65"
Room, outdoor and duct resistance thermometers
Typ: 902520/22-474-1003-1/330,(-40..80°C)
- Measuring range from -40 to +80°C (Measuring range only in conjunction with transmitter)
- Measuring insert 1x temperature sensor Pt100 in two-wire circuit
- Transmitter: 1x analog transmitter, 4-20 mA output (only in conjunction with 1xPt100 tw)(-40 to +80°C)
- Protection class IP65
- Suitable for use in air conditioning technology
Room, outdoor and duct resistance thermometers are designed for air conditioning technology and are mainly used for temperature measurement in rooms, air ducts and outdoors.
Various plastic device versions with different degrees of protection are available for the respective measuring task.
A Pt100 temperature sensor in accordance with DIN EN 60751 class B in a two-wire circuit is installed as standard, but variants with Pt500, Pt1000, Ni1000 and various NTC sensors are also possible. The connection terminals allow optional continuation in three- and four-wire circuits. An integrated measuring transducer can also be selected as an option.
Dimensions
Height | 7 cm |
---|---|
Length | 5,2 cm |
Width | 10 cm |
Weight
Weight | 100 g |
---|
Other attributes
Customs tariff number | 90251900 |
---|---|
Messeinsatz | 1x Pt 100 Zweileiterschaltung |
Messumformer | Messumformer analog Ausgang 4-20 mA |
Temperaturbereich | -40°C ... +80°C |
Toleranz / Klasse | Klasse B |
Connection housing | Plastic housing PC (basic type 902520/11, material PP), protection class IP20 to IP65, basic type 902524/25 protection class IP54 and IP65 |
Protective tube | stainless steel 1.4571; Ø 6 mm |
Measuring insert | Pt100 temperature sensor, DIN EN 60751, class B, two-wire circuit Pt1000 temperature sensor, DIN EN 60751, class B, two-wire circuit, For more information, see order details |
Transmitter | Analog transmitter, output signal 4 to 20 mA or 0 to 10 V |
| Output 4 to 20 mA | Output 0 to 10 V |
Input Measurement Input Sensor Current Measurement Rate |
Pt100 (DIN EN 60751) 0.5 mA Continuous measurement, as analogue signal path |
Pt100 (DIN EN 60751) 0,5 mA Continuous measurement, as analogue signal path |
Measuring Circuit Monitoring Measuring Range Undershooting Measuring Range Exceedance
Sensor short circuit Sensor and cable breakage |
decreasing to 3.6 mA increasing to 22 to < 28 mA (typically 24 mA) 3.6 mA 22 to < 28 mA (typically 24 mA) |
0 V increasing to 11 to < 14 V (typically 12 V) 0 V 11 to < 14 V (typically 12 V) |
Exit |
|
|
Output | eingeprägter Gleichstrom 4 bis 20 mA | Gleichspannung 0 bis 10 V |
Transmission Behavior | temperaturlinear | temperaturlinear |
Transmission Accuracy | ± 0,1 % | ± 0,2 % |
Damping of ripple | 37 dB | 40 dB |
of a power supply 24 V, |
|
|
Amplitude 10 V/50 Hz, Bürde 470 /Last 10 M |
|
|
Burden (Rb) | Rb = (Ub - 7,5 V) ÷ 22 mA | - |
Burden Influence | ± 0,02 %/100 a | - |
Load/Lasteinfluss | - | 10 k/ ± 0,1 % |
Setting time in case of temperature change | 10 ms | 10 ms |
Matching conditions | DC 24 V/ca. 22 °C | DC 24 V/ca. 22 °C |
Matching accuracy | ± 0,2 %a,b or ± 0,2 K | ± 0,2 %a,b or ± 0,2 K |
Overall Accuracy Sensor/Adjustment | ± 0,4 K (typical) for 20 °C/24 V | ± 0,4 K (typical) for 20 °C/24 V |
| Power supply | Power supply |
Power supply Power supply (Ub) Reverse polarity protection Power supply influence |
DC 7,5 until 30 V and ± 0,01 %/V Deviation from 24 Va |
DC 15 until 30 V and ± 0,01 %/V Deviation from 24 Va |
Environmental impact |
|
|
Temperature range | -40 until + 85 °C | -40 until + 85 °C |
Storage temperature range | -40 until +100 °C | -40 until +100 °C |
Temperature | ± 0,01 %/K Deviation from 22 °Ca | ± 0,01 %/K Deviation of 22 °Ca |
Climatic resistance similar to DIN EN 60654 Class C1 | relative humidity 95 % annual average without condensation | relative humidity 95 % annual average without condensation |
EMC Emission Resistance/Immunity | EN 61326 Class B, Industrial Requirements | EN 61326 Class B, Industrial Requirements |