Measurement parameters:
Water hardness | Carbonate hardness | p-value | minus m-value
Measuring range:
0.89 - 448 ppm | 8.9 - 358 ppm | 0.1 – 15 mmol/l | 0.05 – 0.5 mmol/l
Application:
Universal for water treatment plants | Dosing a reducing agent at high concentration of oxidising agent
Menu navigation:
German, English
Our comment:
Best choice for control of water quality in areas where measurement errors can arise due to oxidizing agents!
Would you like to register the warranty for this product?
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Offer as PDF
Login for addressed offer The entered number of pieces will be taken into account when creating the offer.
Login for addressed offer The entered number of pieces will be taken into account when creating the offer.
Product information "Testomat 2000® Antox"
The Testomat 2000® Antox is a special version of the Testomat 2000® with an additional pump for dosing a reducing agent. By adding this reducing agent in the already filled measuring chamber, the disturbance of oxidizing agents is eliminated. The further analysis of the water hardness measurement is then continued according to the standard.
Performance profile:
- Menu-driven operation and programming via plain text display
- Determinal measuring ranges of residual hardness, total hardness, carbonate hardness, minus m - value, p - value by indicator selection
- Selection of hardness units in ° dH, ° f, ppm CaCO3 or mmol / l
- High-precision titration using a piston dosing pump
- Analysis initiation:
- Automatic interval operation (Interval pause adjustable from 0–99 min)
- External control
- Dynamic (fatigue-dependent interval operation)
- Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes)and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- Internal error documentation
- Programmable service address
- Indicator quantities: 100 and 500 ml bottles
- Programmable maintenance interval for maintenance request
Weight
Weight | 10,5 kg |
---|
Other attributes
Country of origin | DE |
---|---|
Customs tariff number | 90278990 |
Technical specifications
Power consumption: | max. 30 VA |
Protection class: | I |
Protection: | IP 65 |
Ambient temperature: | 10–45 °C |
Water temperature: | 10–40 °C |
Dimensions (W x H x D): | 380 x 480 x 280 mm |
Operating pressure: | 0,3–8 bar |
Weight: | approx. 9,5 kg |
Menu Language: | German, English |
Measuring range: | Water hardness: 0.89 - 448 ppm | Carbonate hardness: 8.9 - 358 ppm | p-value: 0.1 – 15 mmol/l | minus m-value: 0.05 – 0.5 mmol/l |