€6,732.00*
Content:
1 Stück
Please log in to receive your conditions.
Available in 70 days, delivery time 1-2 days
Product number:
100216
GTIN:
4068823010848
Measurement parameters:
Water hardness | Carbonate hardness | p-value | minus m-value
Measuring range:
0.89 - 448 ppm | 8.9 - 358 ppm | 0.1 – 15 mmol/l | 0.05 – 0.5 mmol/l
Application:
Universal for water treatment plants | With additional calibration function
Menu navigation:
German, English, French, Italian, Dutch
Our comment:
Additional calibration function!
Would you like to register the warranty for this product?
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Offer as PDF
Login for addressed offer The entered number of pieces will be taken into account when creating the offer.
Login for addressed offer The entered number of pieces will be taken into account when creating the offer.
Leasing Example
140,63 €
non-binding
Monthly Rate (60 Rates)
Prices excl. VAT
Further rate options by choosing the payment method Leasing
Product information "Testomat 2000® CAL"
The Testomat 2000® CAL with additional calibration function automatically determines water hardness, carbonate hardness, the p-value or the minus m-value via titration. It is ideal for monitoring the water quality of water treatment, water blending and drinkingwater plants, monitoring and quality control of softening plants with and without blending (Testomat 2000® reagent TH 2005–2250), monitoring and quality control of decarbonisation plants (Testomat 2000® reagent TC 2050–2100), measuring acidic values in decarbonisation and desalination plants (Testomat 2000® reagent TM 2005) and monitoring of steam boilers (Testomat 2000® reagent TP 2100)
Performance profile:
- Menu-driven operation and programming via plain text display
- Determinal measuring ranges of residual hardness, total hardness, carbonate hardness, minus m-value, p-value by indicator selection
- Selection of hardness units in ° dH, ° f, ppm CaCO3 or mmol / l
- High-precision titration using a piston dosing pump
- Analysis initiation:
- Automatic interval operation (Interval pause adjustable from 0–99 min)
- External control
- Dynamic (fatigue-dependent interval operation)
- Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- Internal error documentation
- Programmable service address
- Indicator quantities: 100 and 500 ml bottles
- Programmable maintenance interval for maintenance request
Weight
Weight | 10 kg |
---|
Other attributes
Country of origin | DE |
---|---|
Customs tariff number | 90278990 |
Technical specifications
Power consumption: | max. 30 VA |
Protection class: | I |
Protection: | IP 65 |
Ambient temperature: | 10–45 °C |
Water temperature: | 10–40 °C |
Dimensions (W x H x D): | 380 x 480 x 280 mm |
Operating pressure: | 0,3–8 bar |
Weight: | approx. 9,5 kg |
Menu Language: | German, English, French, Italian, Dutch |
Measuring range: | Water hardness: 0.89 - 448 ppm | Carbonate hardness: 8.9 - 358 ppm | p-value: 0.1 – 15 mmol/l | minus m-value: 0.05 – 0.5 mmol/l |