Price on request
Product number:
100500
GTIN:
4068823013832
Measurement parameters:
chlorine dioxide ClO2
Measuring range:
0.00-1.88 mg/l and 1.9-4.7 mg/l
Application:
Monitoring of chlorine dioxide ClO2
Menu navigation:
German, English, French
Would you like to register the warranty for this product?
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Register warranty here
A warranty extension is possible after taking out an indicator subscription.
Product information "Testomat 2000® ClO2"
The Testomat 2000® ClO2 process photometer is an online analyzer for monitoring the content of chlorine dioxide (ClO2) in the range of 0–4,7 mg/l (ppm)
Performance profile:
- Analysis via automatic dosing of two reagents
- The analysis result is displayed after a reaction time of approx. 1 min. (Measuring time without rinsing time)
- 2 independently programmable limit contacts for monitoring and control tasks
- Menu-driven operation and programming via plain text display
- High-precision titration using a piston dosing pump
- Analysis initiation:
- Automatic interval operation (Interval pause adjustable from 0–99 min)
- External control
- Dynamic (fatigue-dependent interval operation)
- Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- programmable service address
- Internal error documentation
- Programmable maintenance interval for maintenance request
Other attributes
Country of origin | EU |
---|---|
Customs tariff number | 90278990 |
Technical specifications
Power consumption: | max. 30 VA |
Protection class: | I |
Protection: | IP 65 |
Ambient temperature: | 10–45 °C |
Water temperature: | 10-40 °C |
Dimensions (W x H x D): | 380 x 480 x 280 mm |
Operating pressure: | 0,3–8 bar |
Weight: | approx. 9,5 kg |
Menu Language: | German, English, French |
Measuring range: | 0.00-1.88 mg/l and 1.9-4.7 mg/l |