Skip to main content

Testomat® 808 SIO2

Product number: 100662

Product information "Testomat® 808 SIO2"

The Testomat® 808 SIO02 is a device that was developed for measuring and monitoring deionised water (fully demineralised water) in medical facilities. 
The central importance of silicate monitoring 
Silicate measurement using the Testomat® 808 SIO02 is an essential aspect of compliance with a defined limit value. Precise silicate monitoring is particularly important in process water treatment in hospitals in order to prevent deposits on devices and medical devices (MD) and to ensure the effectiveness of the water treatment systems. You can choose between 10 limit values in the range from 0.3 to 1.2 ppm. By regularly monitoring the silicate content and compliance with the limit values, potential problems can be identified at an early stage and suitable measures can be taken to optimise water treatment.

Dimensions

Height 31,4 cm
Length 13,8 cm
Width 36,4 cm

Weight

Weight 5,2 kg

Other attributes

Country of origin DE
Customs tariff number 90278990
Technical data
  • Mains connection: depending on device version: 24V / 115V / 230V, 50 - 60 Hz
  • Device fuse protection: 230 - 240 V: T0.1 A; 115 V: T0.2 A; 24 V: T0.8 A
  • Mains fuse protection for consumers: max. 4 A (N , L)
  • Power consumption: max. 16 VA, without external load
  • Protection class: I Protection class: IP 54
  • Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1
  • Ambient temperature: 10 - 40 °C
  • Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for the output of status and error messages, max. load 500 Ohm
  • Relay contact load: 230V / 4A AC resistive load
  • Weight: 4350 g
Water connection
  • Operating pressure: Depending on device version
  • 0.3 - 1 bar / 0.3 x 105 to 1 x 105 Pa
  • 1 - 4 bar / 1 x 105 to 4 x 105 Pa
  • From 4 to 8 bar, a pressure reducer must be used (special accessory)
  • Water inlet: Opaque hose with 6 mm outer diameter/4 mm inner diameter
  • Water outlet: Hose with 6 mm outer diameter/4 mm inner diameter
  • Water temperature: 10 to 40 °C

We recommend our reagents, as well as the following accessories

Testomat® 808 SIO2 Reagent set, reagents A 100 ml & B 100 ml
Product number: 140808
100 ml-bottles of original Testomat® 808 SIO2 Reagent Set A+B for monitoring the level of SiO2 in the water. The bottles of original Testomat® 808 SIO2 Reagent can be used immediately in the instrument without cumbersome filling and refilling. Measuring range: Silica SiO2: 0.3 - 1.2 ppm  

€85.00*
Testomat® 808 SIO2 Reagent A - 500 ml
Product number: 141808
500 ml-bottle of original Testomat® 808 SIO2 Reagent A for monitoring the level of SiO2 in the water. The bottles of original Testomat® 808 SIO2 Reagent can be used immediately in the instrument without cumbersome filling and refilling.   Measuring range: Silica SiO2: 0.3 - 1.2 ppm  

€154.00*
Testomat® 808 SIO2 Reagent B - 500ml
Product number: 141809
500 ml-bottle of original Testomat® 808 SIO2 Reagent B for monitoring the level of SiO2 in the water. The bottles of original Testomat® 808 SIO2 Reagent can be used immediately in the instrument without cumbersome filling and refilling.   Measuring range: Silica SiO2: 0.3 - 1.2 ppm  

€154.00*
Adapter set for Testomat® 808
Product number: 37610
In this adapter set are all parts included, that you need to install a Testomat® 808.   Connection set consisting of: plastic hose 6/4 x 1.5 m length   reduction 10 t0 6 mm   shutoff 3/8" to 6 mm

€104.40*
Service set Professional for 2-year maintenance Testomat® 808 SiO2
Product number: 890612
Service set contains:2 x sight glasses 30 x 33 x flat gasket 24 x 21 x insert for screw cap & suction tube 100 ml bottle1 x insert for screw cap & suction tube 500 ml bottle1 x double pump head conversion kit1 x gear motor2 x Hose connector1 x Cleaning brush set Important: After replacement, the pump must be recalibrated in Testomat® 808 SiO2.The adjustment is available from firmware version T808_83-003. Carry out a firmware update for older versions before performing the adjustment! (Serial number less than 253.061)For this calibration, it is important that you use water with a silicate content of 0 ppm, otherwise the zero point calibration will be incorrect and lead to false measurement results.To ensure this, please use the silicate filter cartridge specially developed for this process for the T808 SiO2, article 270344.(For additional information and instructions on the equipment you need for the software update, please refer to the operating instructions. The instructions "Replacing the pump head on Testomat® 808 SiO2" are also available)

€1,093.40*
Annual service set professional Testomat® 808 SiO2
Product number: 890611
Contains:  2 x Sight glass windows 30x33 x Flat gasket 24x2 EPDM 601 x Bottle insert for screw cap and push-fit suction tube, 100 ml bottle1 x Bottle insert for screw cap and push-fit suction tube, 500 ml bottle1 x brushes set

€169.80*
Repair- and service case Testomat® 808 SiO2
Product number: 270343
Contents: 1 x 37570 Solenoid valve complete Testomat® 808 2 x 37580 Insert with screw cap and suction tube 100ml bottle 1 x 37859 double pump head Testomat® 808     1 x 31972 SUB-D null modem cable, 9-pin, 2 x socket, 3m     4 x 37579 Insert with screw cap and suction tube 500ml bottle     2 x 40050 Magnetic stirring core 4 x 37538 Tube adapter Testomat® 808     4 x 40157 Angle screw-in connector, G1/8"-6     2 x 37543 Pipe, PE, D=6 d=4 L=53mm     2 x 37544 Pipe, PE, D=6 d=4 L=140mm     6 x 40170 Viewing window 30x3     6 x 31595 Fuse solder-in, TR5, T0,1A     6 x 31592 Fuse solder-in, TR5, T1,0A     6 x 31584 Fuse, solder-in, TR5, T0.2A 6 x 31585 fuse, T0,315A     8 x 33797 O-ring 3,68 x 1,78     8 x 31666 Fuse G-, 5x20mm, T4A     8 x 33777 Flat gasket 24x2 EPDM 70     4 x 33253 Screw M3x40, A2, DIN 965     8 x 33246 pan-head screw M3x12 DIN 7985H A2     8 x 11264 O-ring 4.5 x 1.5     8 x 11245 O-ring 1,78x1,78     1 x 32286 Adapter USB > Serial -length 1.1 meter     1 x 32287 Cleaning brush set     1 x 37498 USB stick with software tool for T808 and maintenance manual     2 x 37621 Dosing needle Testomat® 808 completeTranslated with www.DeepL.com/Translator (free version)